For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Product Information
Phase Detail
Zebron ZB-XLB-HT Inferno GC Columns<br class="t-last-br" />
Zebron ZB-XLB-HT Inferno is a non-metal si-arylene GC column stable to 400 ˚C

Zebron ZB-XLB-HT Inferno GC Columns

High Temp Stability, Low Bleed

  • Rugged, non-metal si-arylene GC column stable to 400 ˚C
  • Robust column for high temperature bake outs and analysis, such as high molecular weight compounds
  • Provides unique selectivity for conformational analyses
  • Longer lifetime with high temperature, polyimide coated, fused silica tubing
  • Low activity, provides good peak shape for acidic and basic samples
  • Good tool for general screening to identify unknown samples

Recommended Use
Herbicides/Insecticides, PCBs, Pesticides, Unknown Samples



Recommended Use
Used confirmation of pesticides, PCB, or other environmental samples, robust column performance for high temperature bakeouts

Upgrade to Zebron from these similar* phases:

Aroclor 1242: DIN Method 51527

Column:
Zebron ZB-XLB-HT Inferno
Dimension:
30 meter x 0.25 mm x 0.25 µm
Part No.:
7HG-G024-11
Injection:
Split 2:1 @ 250 °C, 1 µL, pressure pulse @ 40 psi for first 0.25 min
Carrier:
Helium @ 1.5 mL/min (constant flow)
Oven Program:
50 °C for 0.5 min to 210 °C @ 40 °C/min for 3 min to 230 °C @ 30 °C/min for 5 min to 250 °C @ 30 °C/min for 5 min to 320 °C
@ 40 °C/min for 2 min
Detection:
ECD @ 350 °C
Sample:
Total concentration of aroclors was 90 ppm in isooctane
1. TCMX
2. BZ# 31
3. BZ# 28
4. BZ# 52
5. BZ# 101
6. BZ# 77
7. BZ# 123
8. BZ# 118
9. BZ# 153
10. BZ# 114
11. BZ# 105
12. BZ# 138
13. BZ# 126
14. BZ# 167
15. BZ# 156
16. BZ# 180
17. BZ# 157
18. BZ# 170
19. BZ# 169
20. BZ# 189
21. DCB

Melamine and Cyanuric Acid by GC-MS

Columns:
Zebron ZB-XLB-HT Inferno
Dimension:
15 meter x 0.25 mm x 0.25 µm
Part No.:
7EG-G024-11
Injection:
On-Column @ 103 °C, 1 µL
Carrier Gas:
Helium @ 1.4 mL/min (constant flow)
Oven Program:
100 °C for 0.5 min to 320 °C @ 25 °C/min
Detector:
MSD @ 325 °C
Sample:
Analytes are 200 ng / 100 µL in BSTFA / Pyridine (1:1)
1. Cyanuric Acid 13C3 (IS)
2. Cyanuric Acid
3. Melamine 13C3 15N3 (IS)
4. Melamine

Good Results for Difficult Samples

Explosives by GC-MS

Columns:
Zebron ZB-XLB-HT Inferno
Dimension:
15 meter x 0.25 mm x 0.25 µm
Part No.:
7EG-G024-11
Injection:
On-Column @ 73 °C, 0.5 µL
Carrier Gas:
Helium @ 1.4 mL/min (constant flow)
Oven Program:
70 °C for 1 min to 140 °C @ 25 °C/min for 4 min to 280 °C @ 25 °C/min
Detector:
MSD @ 325 °C
Sample:
Analytes are 10 ppm in dichloromethane
1. Nitrobenzene
2. 2-Nitrotoluene
3. 3-Nitrotoluene
4. 4-Nitrotoluene
5. Nitroglycerin
6. 2,6-Dinitrotoluene (2,6-DNT)
7. 1,3-Dinitrobenzene (1,3-DNB)
8. 2,4-Dinitrotoluene
9. 2,4,6-Trinitrotoluene (2,4,6-TNT)
10. PETN
11. 1,3,5-Trinitrobenzene (1,3,5-TNB
12. RDX
13. 4-Amino-2,6-dinitrotoluene
14. 3,5-Nitroaniline
15. 2-Amino-4,6-dinitrotoluene
16. Tetryl