A weak cation-exchange functionalized polymeric sorbent that allows for complete retention of basic compounds with a pKa greater than 8, making 100% organic wash conditions possible. A 100% organic wash ensures that the maximum amount of interferences are removed from the target compound.
Brand Strata-X |
Recommended Use Extraction of basic compounds, particularly strong bases |
Weak Cation-Exchange
Ï€-Ï€ Bonding
Hydrophobic Interaction
96-Well-Plate
96-Well-Plate
1, 3, and 6mL Tubes
Giga Tubes (12-150mL Tubes)
On-line Cartridges
Column:
1 mL Methanol
|
Equilibrate:
1 mL Water, pH 6-7
|
Load:
Diluted Sample, pH 6-7
|
Wash 1:
1 mL Water, pH 6-7
|
Wash 2:
1 mL Methanol (collect this fraction to analyze Neutrals/Acids)
|
Elute Any Bases:
2x 500 µL 5 % Formic Acid in Methanol
|
Elute Weak Bases:
2x 500 µL 5 % NH4OH in Methanol
|
Based on 30 mg/1 mL sorbent mass. The above is a convenient starting point for SPE method development, further optimization may be required to tailor the method for your specific extraction needs.