Carbon Load null % |
Particle Size 8 |
Recommended Use Monosaccharides, sugar alcohols, and class separation of di-, tri-, and tetra- saccharides |
Phase | Description | Applications | Bio-Rad® Aminex® | Supelco® Supelcogel™ | Waters® Sugar-Pak™ | Transgenomic® CARBOSep™ | Sepax® Carbomi® |
---|---|---|---|---|---|---|---|
RCM-Monosaccharide (USP L19 packing)* | 8 % cross-linked CALCIUM | – Monosaccharides and sugar alcohols, including sorbitol and mannitol from sweeteners and corn and cane sugars – Class separation of di-, tri-, and tetra- saccharides |
HPX-87C 125-0095 |
Supelcogel Ca | Sugar-Pak 1 | CARBOSep CHO-820 | Carbomix Ca |
Columns:
Rezex RCM-Monosaccharide
Aminex HPX-87C |
|
Dimension:
300 x 7.8 mm
|
|
Mobile Phase:
Water
|
|
Flow Rate:
0.6 mL/min
|
|
Temperature:
80 °C
|
|
Detection:
ELSD
|
|
Sample:
1. Melezitose
2. Maltose 3. Glucose 4. Mannose 5. Fructose 6. Ribitol |
Columns:
Rezex RCM-Monosaccharide
Aminex HPX-87C |
|
Dimension:
300 x 7.8 mm
|
|
Mobile Phase:
Water
|
|
Flow Rate:
0.6 mL/min
|
|
Temperature:
80 °C
|
|
Detection:
ELSD
|
|
Sample:
1. Sucrose
2. Glucose 3. Fructose |
Columns:
Rezex RCM-Monosaccharide
Bio-Rad® Aminex® HPX-87C |
|
Dimension:
300 x 7.8 mm
|
|
Mobile Phase:
Water
|
|
Flow Rate:
0.6 mL/min
|
|
Temperature:
80 °C
|
|
Detection:
ELSD
|
|
Sample:
1. Sucrose
2. Glucose 3. Galactose 4. Fructose 5. Sorbitol |
Column:
Rezex RCM-Monosaccharide
|
|
Dimension:
300 x 7.8 mm
|
|
Part No.:
00H-0131-K0
|
|
Mobile Phase:
Water
|
|
Flow Rate:
0.6 mL/min
|
Temperature:
75 °C
|
|
Detection:
RI
|
|
Sample:
1. Sucrose
2. Glucose 3. Fructose |
Column:
Rezex RCM-Monosaccharide
|
|
Dimension:
300 x 7.8 mm
|
|
Part No.:
00H-0130-K0
|
|
Mobile Phase:
Water
|
|
Flow Rate:
0.6 mL/min
|
Temperature:
75 °C
|
|
Detection:
RI
|
|
Sample:
1. Dp3
2. Dp2 3. Glucose 4. Fructose |
Column:
Rezex RCM-Monosaccharide
|
|
Dimension:
300 x 7.8 mm
|
|
Part No.:
00H-0130-K0
|
|
Mobile Phase:
Water
|
|
Flow Rate:
0.5 mL/min
|
|
Temperature:
80 °C
|
|
Detection:
ELSD
|
|
Sample:
1. Sucrose
2. Glucose 3. Galactose 4. Fructose |
Column:
Rezex RCM-Monosaccharide
|
|
Dimension:
300 x 7.8 mm
|
|
Part No.:
00H-0130-K0
|
|
Mobile Phase:
Water
|
|
Flow Rate:
0.6 mL/min
|
Temperature:
85 °C
|
|
Detection:
RI
|
|
Sample:
1. Melezitose
2. Maltose 3. Glucose 4. Mannose 5. Fructose 6. Ribitol |
Column:
Rezex RCM-Monosaccharide
|
|
Dimension:
300 x 7.8 mm
|
|
Part No.:
00H-0130-K0
|
|
Mobile Phase:
Water
|
|
Flow Rate:
0.5 mL/min
|
|
Temperature:
60 °C
|
|
Detection:
RI
|
|
Sample:
1. Glycerol
2. Methoxypolyethylene Glycol 3. Triethylene Glycol 4. Sorbitol 5. Urea |
Compound | App ID No. |
---|---|
Fructose | 5504, 5505, 14975, 14981, 14977, 14987 |
Galactose | 14977, 14987 |
Glucose | 5504, 5505, 14975, 14977, 14979, 14981, 14987, 14983 |
Glycerol | 5518 |
Maltose | 5504, 14975 |
Mannose | 5504, 14975 |
Melezitose | 5504, 14975 |
Compound | App ID No. |
---|---|
Methoxypolyethylene Glycol | 5518 |
Ribitol | 5504, 14975 |
Sorbitol | 5518, 14977 |
Sucrose | 5505, 14981, 14977, 14987 |
Tartaric Acid | 5504 |
Triethylene Glycol | 5518 |
Urea | 5518 |
Rezex RCM-Monosaccharide |   |
---|---|
Part No. | 00H-0130-K0 |
Ionic Form | Calcium |
Standard Dimensions | 300 x 7.8 mm |
Matrix | Sulfonated Styrene Divinyl Benzene |
Cross Linking | 8 % |
Particle Size (μm) | 8 |
Min. Efficiency (p/m) based on last peak | 35,000 |
Typical Pressure (psi @ Testing Flow Rate) | 260 |
Max. Pressure (psi @ Max Flow Rate) | 1,000 |
Max. Flow Rate (mL/min) | 1.0 |
Max. Temperature (°C) | 85 |
Typical Mobile Phase | Water |
pH Range | Neutral |
Guard Column Part | 03B-0130-K0 |
Cleaning, Regeneration, and Storage   | |
---|---|
Organic Modifiers (Max) | 5 % Methanol, IPA, EtOH |
Inorganic Modifiers | 5 % CaSO4, Ca(NO3)2, CaCl2 |
Avoid | Acids, Bases, Non-Calcium Salts or Metal Ions, > 30 % Acetonitrile |
Cleaning Solvent | 100 % Water |
Flow Rate(mL/min) | 0.4 |
Temperature (°C) | 85 |
Duration (hrs) | 12 |
Regeneration Solvent | 0.1 M Ca(NO3)2 |
Flow Rate (mL/min) | 0.2 |
Duration (hrs) | 4-16 |
Ship/Storage Solvent | Water |