Controlled pore size processing and a proprietary polymerically bonded stationary phase were developed to ensure excellent resolution between priority polycyclic aromatic hydrocarbons (PAHs). Combined with core shell particle technology, incredibly high efficiency and sensitivity at comfortable LC pressures is very achievable.
*Kinetex analytical columns are compatible with all LC system with standard LC column end-fittings threaded in 10-32. Please contact us at Phenomenex.com/chat for any questions and information about fitting compatibility and any other column internal diameters.
Carbon Load null % |
Particle Size 3.5 |
Recommended Use Polymerically bonded C18 phase specifically developed for the separation of EU and EPA priority PAHs. |
Column:
Kinetex 3.5 µm PAH
|
|
Dimension:
100 x 4.6 mm
|
|
Part No.:
00D-4764-E0 |
|
Mobile Phase:
A: Water B: Acetonitrile |
|
Gradient:
Time (min)
0
7 8 9 12 | % B
50 100 100 50 50 |
Flow Rate:
1.2 mL/min |
|
Backpressure:
136 Bar |
|
Temperature:
35 ËšC |
|
Detection:
UV @ 292 nm |
|
Samples:
1. Naphthalene 2. Acenaphthylene 3. Acenaphthene 4. Fluorene 5. Phenanthrene 6. Anthracene 7. Fluoranthene 8. Pyrene 9. Benzo[a]anthracene 10. Chrysene 11. Benzo[b]fluoranthene 12. Benzo[k]fluoranthene 13. Benzo[a]pyrene 14. Dibenz[a,h]anthracene 15. Benzo[g,h,i]perylene 16. Indeno[1,2,3-cd]pyrene |