FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.


GC Technical Tip

Level: Basic

Gas Chromatography (GC) Column Installation - Detector Installation

When installing your capillary GC column, making accurate connections is essential to ensure trouble free operation. In this technical tip we provide advice on the best way to achieve this.

Note: For users with sensitive detectors such as MS and ECD, column conditioning steps should be performed before installing the column to prevent contamination and frequent maintenance of the detector.

  1. Place the column nut and ferrule past the end of the column and cut a centimeter or two off the end of the column. Be sure that the ferrule is the right size and pointing in the correct direction. Inspect the cut with a magnifier and ensure that the cut is square and smooth. (See figure below.) Recut if needed.
  2. Insert the outlet end of the column into the detector exactly the distance prescribed in the instrument manual. Distances will vary between detectors. Tighten the ferrule nut finger-tight then 1/2 turn with a wrench. If the column can still be moved, tighten another 1/4 turn until the column is secure.
  3. Inspect the column connections for leaks using an electronic leak detector. Leaks at the inlet end may introduce oxygen to the column that will result in increased column bleed and damage to the column phase.
Correctly and Incorrectly Cut GC Capillary

cut-gc-capillary

Correct          vs          Incorrect

Cool-Lock Nut for Agilent® and Shimadzu® Systems

NEED SOME HELP? Find Your GC Liner Online!

  • Avoid burning your fingers – cools with the oven
  • Increased reproducibility – locks insertion depth before installation
  • No need for wrench with hand-tightened connections
  • No more Wite-Out, Tipp-Ex, or septa pieces!

Cool-Lock Nut


Zebron GC liner

NEED SOME HELP? Find Your GC Liner Online!

  • Application
  • Injection Type
  • GC System
  • Your Current Liner Part Number

zliner_200